Know what pre-use checks confirm
A detector needs a working gas response and alarms before it is issued for use. OSHA's testing bulletin incorporates ISEA's recommendation for a bump test or calibration check before each day's use, following the manufacturer's instructions.
The bump test or function check does not provide a measure of the instrument's accuracy.
A bump test checks gas access and alarm function. A calibration check compares the reading with a known test-gas concentration against the manufacturer's tolerance. Full calibration adjusts the response. Recording which check was performed makes the result useful to the next person responsible for the instrument.
The bulletin calls for full calibration after a failed bump test or calibration check. An instrument that fails full calibration should be removed from service. The specified test gas, regulator, tubing and adapter belong in the preparation for these checks.
Alarm recognition needs attention during equipment selection and training. Work clothing, hearing protection and site noise can affect how the wearer notices a warning. Assess the available signals in that context and use the site's procedure for the response.